13-2
The signal oriented measurement commands are used to acquire readings. You can use
these high level instructions to control the measurement process. These commands are
summarized in
klqb
Table 13-1
Signal oriented measurement command summary
Command
CONFigure:<function> [<rang>], [<res>], [<clist>]
FETCh?
READ?
MEASure[:<function>]? [<rang>], [<res>], [<clist>]
Channel list parameter (7702):
<clist> = (@SCH)
where: S
= Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Note: Only one channel can be specified in the 7702 <clist> for the MEASure? command.
Model 2790 SourceMeter
Table
13-1.
When measurements are performed, the readings are fed to other enabled
operations.
Section C
explains
additional information on using FETCh?, READ?, and MEASure? to acquire
readings.
®
Switch System Reference Manual
"Data flow (remote
Description
Places the Model 2790 in a "one-shot" measurement
mode for the specified function.
Requests the latest reading.
Performs an ABORt, INITiate, and a FETCh?.
Performs an ABORt, CONFigure:<function>, and a
READ?. (See Note)
operation)" and provides