Keithley SourceMeter 2790 Reference Manual page 392

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
Figure 16-2
Connections for Model 2790 AC volts verification
Model 2790
Model 2790 SourceMeter Switch System
RATIO
CH AVG
CONT
OCOMP
MATH O U T P U T
PERIOD SENSOR
ACI
Ω2
Ω4
SHIFT
DCV
ACV
DCI
FREQ
DELAY
HOLD
LIMITS
ON/OFF
TYPE
MONITOR
CH-OFF
LOCAL
TRIG
EX TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
CONFIG
HALT
TEST
LSYNC
GPIB
CLOSE
OPEN
STEP
SCAN
DIGITS RATE
EXIT
Table 16-3
ACV reading limits
ACV
range
100mV
1V
10V
100V
750V
* If the 5725A amplifier is not available, change the 700V @ 50kHz step to 220V @ 50kHz. Reading
limits for 220V @ 50kHz = 219.36 to 220.64V.
®
Switch System Reference Manual
NOTE Amplifier required only
for 700V, 50kHz output.
INPUT HI
OUTPUT HI
SENSE
INPUT
Ω 4 WIRE
HI
350V
1000V
!
PEAK
PEAK
LO
500V
PEAK
INPUTS
TEMP
RANGE
F
F
R
F
CARD
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
RS-232
ENTER
INPUT LO
Applied AC
voltage
100.0000mV
1.000000V
10.00000V
100.0000V
700.000V*
Amplifier (Connect to calibrator)
Shielded
Calibrator (Output AC Voltage)
cable
OUTPUT LO
1kHz reading limits
(1 year, 18°C to 28°C)
99.910 to 100.090mV
0.99910 to 1.00090V
9.9910 to 10.0090V
99.910 to 100.090V
699.36 to 700.64V
50kHz reading limits
(1 year, 18°C to 28°C)
99.830 to 100.170mV
0.99830 to 1.00170V
9.98300 to 10.0170V
99.830 to 100.170V
698.79 to 701.21V
16-11