General bus commands

General commands are those commands, such as DCL, that have the same general mean-
ing regardless of the instrument.
Table 10-1
General bus commands
Command
REN
IFC
LLO
GTL
DCL
SDC
GET
SPE, SPD

REN (remote enable)

The remote enable command is sent to the Model 2790 by the controller to set up the
instrument for remote operation. Generally, the instrument should be placed in the remote
mode before you attempt to program it over the bus. Setting REN true does not place the
instrument in the remote state. You must address the instrument to listen after setting REN
true before it goes into remote.
The Model 2790 must be in remote in order to use the following commands to trigger and
acquire readings:
Goes into effect when next addressed to listen.
Goes into talker and listener idle states.
LOCAL key locked out.
Cancel remote; restore Model 2790 front panel operation.
Cancel remote; restore front panel operation for all devices.
Returns all devices to known conditions.
Returns Model 2790 to known conditions.
Initiates a trigger.
Serial polls the Model 2790.
INITiate and then FETCh?
READ?
MEASure?
Table 10-1
lists the general bus commands.
Effect on Model 2790
10-7