16-36
3.
4.
5.
6.
7.
8.
Table 16-16
Limits for Model 7751 source-voltage resistance verification
Source
current
500V
500V
50V
50V
* Calculate limits based on actual characterized resistance values and Model 7751 resistance accuracy
specifications. See Verification limits.
Set the Model 7751/7753 source voltage to 500V as follows:
a.
Press SHIFT, CARD, then CONFIG.
b.
Select SLOT1: 7751 or SLOT1: 7753, then press ENTER.
c.
Select the V-C submenu, then set the source to 500V.
d.
Press EXIT to return to normal display.
Using the CLOSE key, close the following slot 2 channels: 1, 2, 16, 18, 21, 22,
and 23.
Set the math function to display resistance as follows:
a.
Press SHIFT then MATH.
b.
Select S1VOHMS, then press ENTER.
c.
Press EXIT to return to normal display.
Recalculate reading limits shown in
based on the actual characterized resistance value.
Verify that the resistance reading is within the limits calculated in step 6.
Repeat steps 1 through 7 for the remaining resistances and voltage source settings
as shown in
Table 16-16
Nominal
Nominal reading limits
resistance
(1 year, 18°C to 28°C)
1GΩ
0.98186 to 1.01814GΩ
100MΩ
98.585 to 101.415MΩ
100MΩ
94.048 to 105.952MΩ
10MΩ
9.4747 to 10.5253MΩ
Model 2790 SourceMeter
Table 16-16
(7751) or
Table 16-17
(7753).
®
Switch System Reference Manual
(7751) or
Table 16-17
(7753)
Recalculated limits*