Verifying the current source

Check the Model 7751/7752/7753 current source accuracy by setting the current source to
a specific value and then measuring the current with a DMM.
Follow these steps to verify current source accuracy:
1.
2.
3.
4.
5.
Figure 16-16
Connections for Model 7751/7752/7753 current source verification
HP3458A Digital Multimeter
Connect the Model 7751/7752/7753 current source terminals to the DMM, as
shown in
Figure
16-16.
Install the Model 7751/7752/7753 in Slot 1 of the Model 2790, then turn on the
power and allow the unit to warm up for two hours before proceeding. Be sure the
front panel INPUTS switch is set to the REAR position.
Select the DMM DC amps function and enable autorange.
Set the Model 7751/7752/7753 source current to 50mA as follows:
a.
Press SHIFT, CARD, then CONFIG.
b.
Select SLOT1: 7751, SLOT1: 7752, or SLOT1: 7753, then press ENTER.
c.
Select the I-C submenu, then set the source to 50mA.
d.
Press EXIT to return to normal display.
Verify that the DMM reading is within these limits: 49.96mA to 50.04mA.
Model 2790 SourceMeter
INPUT LO
AMPS
®
Switch System Reference Manual
Model 7751, 7752, or 7753
BANK 1
SOURCE SENSE
SRC
SEN
SRC
SEN
A
B
GUARD
HI
LO
HI
LO
HI
HI
LO
LO
GND
GND
HI
LO
HI
LO
I / V
IN
OUT
LO
HI
HI
LO
LO
/ INTERLOCK
VSOURCE
ISOURCE
GUARD
I / V
SRC
SEN
SRC
SEN
A
B
BANK 3
7751, 7752, or 7753
J107
ISOURCE HI
ISOURCE LO
BANK 2
SRC
SEN
SRC
SEN
A
B
HI
HI
LO
LO
HI
HI
LO
LO
SRC
SEN
SRC
SEN
A
B
BANK 4

Advertisement

Table of Contents
loading

Table of Contents