Verifying AC voltage

Check AC voltage accuracy by applying accurate AC voltages at specific frequencies from
the AC voltage calibrator to the Model 7702 inputs. Verify that the displayed readings fall
within specified ranges.
`^rqflk
Follow these steps to verify AC voltage accuracy:
1.
Figure 16-9
Connections for Model 7702 AC volts verification
CH1
Model 7702
2.
3.
4.
Do not exceed 300V RMS between Model 7702 INPUT H and L termi-
nals or between adjacent channels, or 8 × 10
instrument damage may occur.
Connect the Model 7702 CH1 H and L INPUT terminals to the AC voltage calibrator
as shown in
Figure
16-9.
TE122
TE121
Install the Model 7702 in Slot 1 of the Model 2790, then turn on the power and
allow the unit to warm up for two hours before proceeding. Be sure the front panel
INPUTS switch is set to the REAR position.
Select the AC volts function by pressing the ACV key. Close Channel 1 by press-
ing the CLOSE key and then keying in 101.
Set the Model 2790 for the 100mV range; make sure that REL is disabled.
Model 2790 SourceMeter
Amplifier (Connect to calibrator)
OUTPUT
HI
Shielded
Calibrator (Output AC Voltage)
Cable
OUTPUT
LO
®
Switch System Reference Manual