Keithley SourceMeter 2790 Reference Manual page 470

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter Switch System Reference Manual
Ohms temperature coefficients
For measurements 1MΩ to 10MΩ
TC % of reading = (0.02 + 1/V
TC ohms offset = 200Ω / °C
where: V
Example – Determine temperature coefficient for a V-source setting of 100V:
% of reading
Ohms temperature coefficient (100V) = (0.03% + 200Ω) / °C
For measurements 10MΩ to 100MΩ
TC % of reading = (0.02 + 6/V
TC ohms offset = 200Ω / °C
where: V
Example – Determine temperature coefficient for a V-source setting of 100V:
% of reading
Ohms temperature coefficient (100V) = (0.08% + 200Ω) / °C
For measurements 100MΩ to 1GΩ
TC % of reading = (0.02 + 51/V
TC ohms offset = 200Ω/ °C
where: V
Example – Determine temperature coefficient for a V-source setting of 100V:
% of reading
Ohms temperature coefficient (100V) = (0.53% + 200Ω) / °C
SOURCE
is the voltage output setting in volts
SOURCE
= (0.02 + 1/100) % / °C
= (0.02 + 0.01) % / °C
= 0.03% / °C
SOURCE
is the voltage output setting in volts
SOURCE
= (0.02 + 6/100) % / °C
= (0.02 + 0.06) % / °C
= 0.08% / °C
is the voltage output setting in volts
SOURCE
= (0.02 + 51/100) % / °C
= (0.02 + 0.51) % / °C
= 0.53% / °C
) % / °C
) % / °C
) % / °C
SOURCE
Specifications
A-11