Keithley SourceMeter 2790 Reference Manual page 115

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

3-10
Model 7702 switching module
Connections for the 7702 switching module are shown in
ACV measurements
Ratio and channel average calculations — Ratio calculates the reading ratio of two
channels, while channel average calculates the reading average of two channels. For these
calculations, paired switching channels are used. Primary channels 1 through 20 are paired
to channels 21 through 40 (channel 1 paired to channel 21, channel 2 paired to channel 22,
and so on). As shown in
channel (i.e., 104) and the other source is connected to its paired channel (i.e., 124).
klqb
Figure 3-3
DCV and ACV connections using Model 7702 switching module
A) DCV connections
Model 7702
Switching
Module
B) ACV connections
Model 7702
Switching
Module
C) Ratio and channel average connections (DCV)
DC
Voltage
Source
NOTE The low connections for channels 1 through 20 do not
need to be referenced to the low connections for
channels 21 through 40.
Model 2790 SourceMeter
(Figure
3-3A and B), channels 1 through 40 can be used.
Figure
3-3C, one DC voltage source is connected to a primary
The ratio and channel average calculations are covered in
H
CH 1-40
L
H
CH 1-40
L
H
Model 7702
CH 21-40
Switching
L
CAUTION Maximum input: 300VDC or RMS,
1A switched, 60W, 125VA maximum.
DC Voltage
Source
AC Voltage
Source
H
CH 1-20
Module
L
®
Swich System Reference Manual
Figure
3-3. For basic DCV and
Section
5.
Voltage
Source
DC