Keithley SourceMeter 2790 Reference Manual page 158

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
Table 4-5
Rate and bandwidth commands
Commands
Integration rate commands
[SENSe[1]]
:VOLTage[:DC]:NPLCycles <n> [, <clist>]
:VOLTage[:DC]:APERture <NRf>
[, <clist>]
:VOLTage:AC:NPLCycles <n> [, <clist>]
:VOLTage:AC:APERture <NRf> [, <clist>]
:CURRent[:DC]:NPLCycles <n> [, <clist>]
:CURRent[:DC]:APERture <NRf>
[, <clist>]
:CURRent:AC:NPLCycles <n> [, <clist>]
:CURRent:AC:APERture <NRf> [, <clist>]
:RESistance:NPLCycles <n> [, <clist>]
:RESistance:APERture <NRf> [, <clist>]
:FRESistance:NPLCycles <n> [, <clist>]
:FRESistance:APERture <NRf> [, <clist>]
:TEMPerature:NPLCycles <n> [, <clist>]
:TEMPerature:APERture <NRf> [, <clist>]
Bandwidth commands
[SENSe[1]]
:VOLTage:AC:DETector:BANDwidth
<NRf> [, <clist>]
:CURRent:AC:DETector:BANDwidth
<NRf> [, <clist>]
Channel list parameter:
7702 <clist> = (@SCH)
where: S
= Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
7751/7752/7753 <clist> (DCV, Ω2, Ω4) = (@Mn) (see
Notes:
1. The <clist> parameter is used to configure one or more 7702 channels for a scan. Each channel in the <clist> must be set to the
function specified by the command. If not, a conflict error (-221) will occur. For example, RESistance:NPLCycles 1, (@101) is
only valid if scan channel 101 is set for the ¾2 function.
2. For 60Hz line power, x = 60. For 50Hz line power, x = 50.
3. For 60Hz line power, x = 1.67e-4. For 50Hz line power, x = 2e-4.
4. For 60Hz line power, the default is 16.67msec. For 50Hz line power, the default is 20msec.
5. Commands to set rate for ACV and ACI are only valid if bandwidth is set to 300 (300Hz to 300kHz). See
conflict error," page
4-12, for details.
6. The instrument will actually accept a parameter value up to 10e6, but it will default to 3e5.
7. [:DC] is optional for the commands to set DCV and DCI integration rate.
®
Switch System Reference Manual
1, 7
Description
Optional root command.
Set rate for DCV in PLCs; <n> = 0.01 to x
Set rate for DCV in secs; <NRf> = x to 1
Set rate for ACV in PLCs; <n> = 0.01 to xx
Set rate for ACV in secs; <NRf> = x to 1
Set rate for DCI in PLCs; <n> = 0.01 to x
Set rate for DCI in secs; <NRf> = x to 1
Set rate for ACI in PLCs; <n> = 0.01 to x
Set rate for ACI in secs; <NRf> = x to 1
Set rate for Ω2 in PLCs; <n> = 0.01 to x
Set rate for Ω2 in secs; <NRf> = x to 1
Set rate for Ω4 in PLCs; <n> = 0.01 to x
Set rate for Ω4 in secs; <NRf> = x to 1
Set rate for TEMP in PLCs; <n> = 0.01 to x
Set rate for TEMP in secs; <NRf> = x to 1
Optional root command.
Set AC bandwidth for ACV in Hertz;
6
<NRf> = 3 to 3e5
.
Set AC bandwidth for ACI in Hertz;
6
<NRf> = 3 to 3e5
.
"Memory patterns," page
4-11
Default
2
.
5.0
3
.
(Note 4)
2,5
. 5.0
3,5
.
(Note 4)
2
.
5.0
3
.
(Note 4)
2,5
.
5.0
3,5
.
(Note 4)
2
.
5.0
3
.
(Note 4)
2
.
5.0
3
.
(Note 4)
2
.
5.0
3
.
(Note 4)
30
30
2-36).
"Rate and bandwidth