Replacing Model 7702 plug-in module amps fuses

t^okfkd
1.
2.
3.
4.
Figure 18-3
Model 7702 amps fuses
CH41 Fuse
®
Switch System Reference Manual
The information in this section is intended only for qualified service
personnel. Do not perform these procedures unless you are qualified to
do so.
Make sure that all plug-in module connections are de-energized and
disconnected before replacing module amps fuses.
Turn off the power and disconnect external connections from the Model 7702.
Open the Model 7702 top cover.
Locate the amps fuses for CH41 and CH42 (see
Remove the circuit board from the bottom plastic housing by removing the two
bottom screws.
Model 7702
TE122
TE121
CH42 Fuse
Figure
18-3).