5-4
Remote programming — rel
Rel commands
The rel commands to set range are listed in
commands follow the table.
klqb
Table 5-1
Rel commands
Commands
Rel commands for DCV
[SENSe[1]]
:VOLTage[:DC]:REFerence <n> [, <clist>]
:VOLTage[:DC]:REFerence:STATe <b>
[, <clist>]
:VOLTage[:DC]:REFerence:ACQuire [, <clist>]
Rel commands for ACV
[SENSe[1]]
:VOLTage:AC:REFerence <n> [, <clist>]
:VOLTage:AC:REFerence:STATe <b> [, <clist>]
:VOLTage:AC:REFerence:ACQuire [, <clist>]
Rel commands for DCI
[SENSe[1]]
:CURRent[:DC]:REFerence <n> [, <clist>]
:CURRent[:DC]:REFerence:STATe <b>
[, <clist>]
:CURRent[:DC]:REFerence:ACQuire [, <clist>]
Rel commands for ACI
[SENSe[1]]
:CURRent:AC:REFerence <n> [, <clist>]
:CURRent:AC:REFerence:STATe <b>
[, <clist>]
:CURRent:AC:REFerence:ACQuire [, <clist>]
Model 2790 SourceMeter
Query commands are not included in
subsystem are provided in
1
®
Table
5-1. Additional information on these
Table
5-1. All commands for the SENSe
Table
15-5.
Description
Optional root command.
Specify rel value; <n> = -1010 to 1010
(V).
Enable/disable rel; <b> = ON or OFF.
Use input signal as rel value.
Optional root command.
Specify rel value; <n> = -757.5 to 757.5
(V).
Enable/disable rel; <b> = ON or OFF.
Use input signal as rel value.
Optional root command.
Specify rel value; <n> = -3.1 to 3.1 (A).
Enable/disable rel; <b> = ON or OFF.
Use input signal as rel value.
Optional root command.
Specify rel value; <n> = -3.1 to 3.1 (A).
Enable/disable rel; <b> = ON or OFF.
Use input signal as rel value.
Switch System Reference Manual
Default
0
OFF
0
OFF
0
OFF
0
OFF

Advertisement