Keithley SourceMeter 2790 Reference Manual page 241

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

7-36
Table 7-2
External trigger scan example
1
Restore defaults:
Restore defaults (SHIFT SETUP > RESTORE: FACT).
2
For front panel operation, proceed to step 3.
For remote programming, clear buffer and disable
buffer auto clear.
3
Configure advanced scan:
(SHIFT CONFIG > ADVANCED):
a
Channel 101:
Select TEMP function.
Configure temperature (SHIFT SENSOR):
Select thermistor sensor (SENS: THRMSTR).
Select thermistor type (TYPE: 5000Ω).
b
Channel 102:
Select Ω2 function.
Select 1MΩ range.
c
Disable (off) channels 103 through 142 (SHIFT CH-OFF).
d
Enable immediate scan (IMM SCAN: Y).
e
Disable timer (TIMER? NO).
f
Set reading count to infinity (RDG CNT: INF).
4
Select external trigger control source:
Press EX TRIG.
5
Enable scan:
Press SCAN.
Front panel operation
Model 2790 SourceMeter
*RST
TRAC:CLE
TRAC:CLE:AUTO OFF
FUNC 'TEMP',(@101)
TEMP:TRAN THER,(@101)
TEMP:THER 5000,(@101)
FUNC 'RES',(@102)
RES:RANG 1e6,(@102)
ROUT:SCAN (@101,102)
ROUT:SCAN:TSO IMM
TRIG:COUN INF
SAMP:COUN 2
TRIG:SOUR EXT
ROUT:SCAN:LSEL INT
INIT
®
Switch System Reference Manual
Remote programming