4-4
Remote programming — range
Range commands
The commands to set range are listed in
mands follow the table.
klqb
Table 4-2
Range commands
Commands
[SENSe[1]]
:VOLTage[:DC]:RANGe[:UPPer] <n> [, <clist>]
:VOLTage[:DC]:RANGe[:UPPer]:AUTO <b>
[, <clist>]
:VOLTage:AC:RANGe[:UPPer] <n> [, <clist>]
:VOLTage:AC:RANGe[:UPPer]:AUTO <b>
[, <clist>]
:CURRent[:DC]:RANGe[:UPPer] <n> [, <clist>]
:CURRent[:DC]:RANGe[:UPPer]:AUTO <b>
[, <clist>]
:CURRent:AC:RANGe[:UPPer] <n> [, <clist>]
:CURRent:AC:RANGe[:UPPer]:AUTO <b>
[, <clist>]
:RESistance:RANGe[:UPPer] <n> [, <clist>]
:RESistance:RANGe[:UPPer]:AUTO <b>
[, <clist>]
:FRESistance:RANGe[:UPPer] <n> [, <clist>]
:FRESistance:RANGe[:UPPer]:AUTO <b>
[, <clist>]
Channel list parameter:
7702 <clist> = (@SCH)
where: S
= Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
7751/7752/7753 <clist> (DCV, Ω2, Ω4) = (@Mn) (see
1. The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the func-
tion specified by the range command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:RANGe 1, (@101) is
only valid if scan channel 101 is set for the ACV function.
2. [:DC] is optional for the commands to set DCV and DCI range.
Model 2790 SourceMeter
Query commands and some optional command words are not included in
Table
4-2. All commands for the SENSe subsystem are provided in
1, 2
®
Table
4-2. Additional information on these com-
Description
Optional root command.
Select DCV range; <n> = 0 to 1010 (V).
Control DCV auto range; <b> = ON or OFF.
Select ACV range; <n> = 0 to 757.5 (V).
Control ACV auto range; <b> = ON or OFF.
Select DCI range; <n> = 0 to 3 (A).
Control DCI auto range; <b> = ON or OFF.
Select ACI range; <n> = 0 to 3 (A).
Control ACI auto range; <b> = ON or OFF.
Select Ω2 range; <n> = 0 to 120e6 (Ω).
Control Ω2 auto range; <b> = ON or OFF.
Select Ω4 range; <n> = 0 to 120e6 (Ω).
Control Ω4 auto range; <b> = ON or OFF.
"Memory patterns," page
Switch System Reference Manual
Table
2-36).
15-5.
Default
1000
ON
750
ON
3
ON
3
ON
120e6
ON
120e6
ON