Model 2790 SourceMeter
Exercise 3 — Closing and opening 7751/7752/7753 channels
(multiple channel operation)
The exercise in
Model 7751/7752/7753 installed in slot 1 of the mainframe.
Table 1-8
Exercise 3 — Close and open 7751/7752/7753 channels (multiple channel operation)
1 Open all channels*:
Press OPEN > display OPEN:ALL > Press OPEN.
2 Select Ω2 function:
Press Ω2.
3 Close channels to connect input channel 101 to DMM Input of 2790:
Press CLOSE > display CLOSE:MULTI > key in 101 > press ENTER >
key in 114 > press ENTER > key in 118 > press ENTER > press EXIT
4 Open all channels*:
Press OPEN > display OPEN:ALL > Press OPEN
*It is a good, safe practice to start and end a switching sequence by opening all channels.
®
Switch System Reference Manual
Table 1-8
demonstrates a sequence to close and open channels of a
Front panel operation
1-31
Command sequence
ROUT:OPEN ALL
FUNC 'RES'
ROUT:MULT:CLOS
(@101,114,118)
ROUT:OPEN ALL