G-10

:STEP5

(:CALibration:PROTected:DC:STEP5)
Purpose
Format
Parameters
Description
Example

:STEP6

(:CALibration:PROTected:DC:STEP6)
Purpose
Format
Parameters
Description
Example

:STEP7

(:CALibration:PROTected:DC:STEP7)
Purpose
Format
Parameters
Description
Example
Model 2790 SourceMeter® Switch System Reference Manual
To program the 100V DC comprehensive calibration step.
:cal:prot:dc:step5 <Cal_voltage>
<Cal_voltage> = 90 to 110 [V]
:STEP5 programs the 100V DC comprehensive calibration step. The
allowable range of the calibration voltage parameter is from 90 to 110,
but 100 is recommended for best results.
:CAL:PROT:DC:STEP5 100
To program the 1kΩ 4-wire comprehensive calibration step.
:cal:prot:dc:step6 <Cal_resistance>
<Cal_resistance> = 0.9E3 to 1.1E3 [Ω]
:STEP6 programs the 1kΩ 4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
0.9E3 to 1.1E3, but 1E3 is recommended for best results.
:CAL:PROT:DC:STEP6 1E3
To program the 10kΩ 4-wire comprehensive calibration step.
:cal:prot:dc:step7 <Cal_resistance>
<Cal_resistance> = 9E3 to 11E3 [Ω]
:STEP7 programs the 10kΩ 4-wire resistance comprehensive calibra-
tion step. The allowable range of the calibration resistance parameter is
from 9E3 to 11E3, but 10E3 is recommended for best results.
:CAL:PROT:DC:STEP7 10E3
Program 100V step.
Program 1kΩ step.
Program 10kΩ step.