Range; Measurement Ranges And Maximum Readings - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

4-2

Range

The range setting is "remembered" by each measurement function. When you select a
function, the instrument will return to the last range setting for that function.

Measurement ranges and maximum readings

The selected range affects both accuracy of the measurement as well as the maximum
level that can be measured. The measurement ranges and maximum readings for all func-
tions, except FREQ, PERIOD, and TEMP, are listed in
Input values that exceed the maximum readings cause the message "OVERFLOW" to be
displayed.
Table 4-1
Measurement ranges and maximum readings
Function
DCV
ACV
DCI
ACI
Ω2
Ω4
*Offset compensated ohms (OCOMP) can be performed on the 100¾, 1k¾, and 10k¾ ranges.
FREQ and PERIOD — Frequency measurements from 3Hz to 500kHz and period
measurements from 2µs to 333µs can be made on the ACV ranges.
TEMP — There is no range selection for temperature measurements. Temperature
measurements are performed on a single, fixed range. Depending on which type of sensor
is being used, the maximum temperature readings range from –200°C to 630°C.
Appendix
100mV, 1V, 10V, 100V, 1000V
100mV, 1V, 10V, 100V, 750V
20mA, 100mA, 1A, 3A
1A, 3A
100Ω, 1kΩ, 10kΩ, 100kΩ, 1MΩ, 10MΩ, 100MΩ
*
100Ω, 1kΩ, 10kΩ, 100kΩ, 1MΩ, 10MΩ, 100MΩ
A, "Specifications," lists the reading range for each sensor type.
Model 2790 SourceMeter
Ranges
®
Switch System Reference Manual
Table
4-1.
Maximum
reading
±1010V
757.5V
±3.1A
3.1A
120MΩ
120MΩ