2-46
Table 2-5 (cont.)
Command sequence for memory patterns test example
Command sequence
:FORM:ELEM READ
:TRAC:DATA?
Note: Nine values will be returned in a single string with comma separators between the values. The string
will have to be parsed to obtain the individual readings. The inspection process can actually be automated by
assigning limits to the memory locations. Doing so enables the instrument to identify Pass/Fail conditions
during the scan. The instrument can also be configured to generate digital outputs in response to these
conditions.
SAMP:COUN 1
Model 2790 SourceMeter
Send READINGS only when data is requested.
Since this is same as specified in the Initial Setup,
this command can be left out.
Request data stored in buffer. Enter the readings and
inspect the values – take action as required.
Restore Sample Count to 1 for normal (non-scanning)
operation.
®
Switch System Reference Manual
Description