5. Install the top cover, and then configure the system as described in
Figure 6: A through D relay installation
Cover plate

System configuration

After installing the new relays, you must program the S46L for the new relay configuration using the following
command:
:ROUT:CONF:CPOL <clist>
Where <clist> is defined as:
clist = (@0 | 3-6, 0 | 3-6, 0 | 3-6, 0 | 3-6, 0 | 1, ... 0 | 1)
Relay:
A
B
NOTE
All switch locations must be included in the <clist> parameter whether or not all
locations are populated. The numeric value (1 or 3-6) indicates the number of
throws; a value of 0 indicates that a location is not populated.
For example, if RELAY A and RELAY B are SP6T, RELAY 1 and RELAY 2 are
SPDTs, and all other locations are not populated, the command syntax is:
:ROUT:CONF:CPOL (@6,6,0,0,1,1,0,0)
See Section 3 of the S46/S46T/S46L Instruction Manual for detailed programming information.