Output Trigger — The Model 2790 will send one or more output triggers. The output
trigger is applied to the Trigger Link connector on the rear panel. It can be used to trigger
an external instrument to perform an operation.
The trigger model can be configured to output a trigger after the completion of a series of
measurements or after every measurement. For example, with the sample counter set to 10
and the trigger counter set to one, a trigger will be sent after the 10 measurements are per-
formed. If instead, the trigger counter is set to 10 and the sample counter is set to 1, a trig-
ger will be sent after each measurement.

Triggering commands

Commands for triggering are summarized in
Table 8-2
SCPI commands — triggering
Commands
ABORt
INITiate[:IMMediate]
INITiate:CONTinuous <b>
FETch?
READ?
TRIGger:SOURce <name>
TRIGger:TIMer <n>
TRIGger:COUNt <NRf>
TRIGger:DELay <n>
TRIGger:DELay:AUTO <b>
TRIGger:SIGNal
SAMPle :COUNt <NRf>
[:SENSe[1]
:HOLD:WINDow <NRf>
:HOLD:COUNt <NRf>
:HOLD:STATe <b>
:SYSTem:BEEPer:STATe <b> Enable or disable the beeper.
*RST
Notes:
1. SYSTem:PRESet enables continuous initiation.*RST disables continuous initiation.
2. SYSTem:PRESet sets the trigger count to INF (infinity). *RST sets the count to 1.
Model 2790 SourceMeter
Description
Reset trigger system.
Initiate one trigger cycle.
Enable/disable continuous initiation; ON or OFF.
Request the last reading(s).
Perform an ABORt, INITiate, and a FETch?.
Select control source; IMMediate, TIMer, MANual,
BUS, or EXTernal.
Set timer interval; 0 to 999999.999 (sec).
Set trigger count; 1 to 55000 or INFinity.
Set delay; 0 to 999999.999 (sec).
Enable or disable auto delay.
Loop around control source.
Set sample count; 1 to 55000.
Optional root command for HOLD commands.
Set Hold window in %; <NRf> = 0.01 to 20.
Set Hold count; <NRf> = 2 to 100.
Enable or disable Hold.
Restore *RST defaults (see "Default" column of this
table). Places Model 2790 in the idle state.
®
Switch System Reference Manual
Table
8-2.
Default Ref
a
b
Note 1
c
d
d
IMM
e
0.1
Note 2
0
f
g
1
h
i
1
5
OFF
ON

Advertisement

Table of Contents
loading

Table of Contents