Model 2790 SourceMeter® Switch System Reference Manual

:LOCK

(:CALibration:PROTected:CARD1:LOCK)
Purpose
Format
Parameters
Description
Example
:LOCK?
(:CALibration:PROTected:CARD1:LOCK?)
Purpose
Format
Response
Description
Example

:RCOunt

(:CALibration:PROTected:CARD1:RCOunt)
Purpose
Format
Parameters
Description
Example
To lock out Model 7751/7752/7753 calibration.
:cal:prot:card1:lock
None
The :LOCK command allows you to lock out Model 7751/7752/7753
calibration after completing the procedure. Thus, :LOCK performs the
opposite of enabling calibration with the :CODE command.
:CAL:PROT:CARD1:LOCK
To read Model 7751/7752/7753 calibration lock status.
:cal:prot:card1:lock?
0
Calibration locked.
1
Calibration unlocked.
The :LOCK? query requests status from the Model 2790 on
Model 7751/7752/7753 calibration locked/unlocked state. Calibration
must be enabled sending the :CODE command before calibration can
be performed.
:CAL:PROT:CARD1:LOCK?
To reset card calibration count to 0.
:cal:prot:card1:rco
None
The :RCOunt command resets the card calibration count reported by
:COUNt? to 0.
:CAL:PROT:CARD1:RCO
Lock out card 1 calibration.
Request card 1 cal lock state.
Reset card calibration count.
G-19