Keithley SourceMeter 2790 Reference Manual page 428

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
2.
3.
klqb
®
Switch System Reference Manual
Set the calibrator to output DC volts and turn external sense off.
Perform the steps listed in
calibration step:
Set the calibrator to the indicated value and make sure it is in OPERATE.
Press the ENTER key to calibrate that step.
Wait until the Model 2790 finishes each step. (The unit will display the
CALIBRATING message while calibrating.)
If your calibrator cannot output the values recommended in
left and right arrow keys and the up and down range keys to set the Model 2790
display value to match the calibrator output voltage.
Table 17-3
DC volts calibration summary
Calibration
step
+10V
-10V
100V
Table 17-3
to complete DC volts calibration. For each
Calibrator
voltage
Allowable range
+10.00000V
+9V to +11V
-10.00000V
-9V to -11V
+100.0000V
+90V to +110V
17-9
Table
17-3, use the