Keithley SourceMeter 2790 Reference Manual page 390

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
2.
3.
4.
5.
®
Switch System Reference Manual
Select the DC volts function by pressing the DCV key and set the Model 2790 to
the 100mV range.
Set the calibrator output to 0.00000mV DC and allow the reading to settle.
Enable the Model 2790 REL mode. Leave REL enabled for the remainder of the
DC volts verification tests.
Source positive and negative and full-scale voltages for each of the ranges listed in
Table
16-2. For each voltage setting, be sure that the reading is within stated limits.
Table 16-2
DCV reading limits
DCU
Applied
Range
DC voltage*
100mV
100.0000mV
1V
1.000000V
10V
10.00000V
100V
100.0000V
1000V
1000.000V
*Source positive and negative values for each range.
Reading limits
(1 year, 18° to 28°C)
99.9900 to 100.0100mV
0.999963 to 1.000037V
9.99965 to 10.00035V
99.9936 to 100.0064V
999.940 to 1000.059V
16-9