Model 2790 SourceMeter
Table 17-11
AC voltage calibration programming steps
Calibration step
10mV AC at 1kHz
100mV AC at 1kHz
100mV AC at 50kHz
1VAC at 1kHz
1VAC at 50kHz
10VAC at 1kHz
10VAC at 50kHz
100VAC at 1kHz
100VAC at 50kHz
700VAC at 1kHz
AC current calibration
Follow these steps for AC current calibration:
1.
2.
Table 17-12
AC current calibration programming steps
Calibration step
100mA at 1kHz
1A at 1kHz
2A at 1kHz
®
Switch System Reference Manual
Calibrator voltage,
10.00000mV, 1kHz
100.0000mV, 1kHZ
100.0000mV, 50kHz
1.000000V, 1kHz
1.000000V, 50kHz
10.00000V, 1kHz
10.00000V, 50kHz
100.0000V, 1kHz
100.0000V, 50kHz
700.000V, 1kHz
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2790
as shown in
Figure
17-3.
Perform the calibration steps summarized in
Set the calibrator to the indicated current and frequency. Make sure the unit is
in operate. (You must use the stated current and frequency.)
Send the indicated programming command.
Wait until the Model 2790 completes each step before continuing.
Calibrator current,
frequency
100.0000mA, 1kHz
1.000000A, 1kHz
2.000000A, 1kHz
frequency
Calibration command
:CAL:PROT:AC:STEP1
:CAL:PROT:AC:STEP2
:CAL:PROT:AC:STEP3
:CAL:PROT:AC:STEP4
:CAL:PROT:AC:STEP5
:CAL:PROT:AC:STEP6
:CAL:PROT:AC:STEP7
:CAL:PROT:AC:STEP8
:CAL:PROT:AC:STEP9
:CAL:PROT:AC:STEP10
Table
Calibration command
:CAL:PROT:AC:STEP11
:CAL:PROT:AC:STEP12
:CAL:PROT:AC:STEP13
17-12. For each step: