Model 2790 SourceMeter® Switch System Reference Manual

:CODE

(:CALibration:PROTected:CODE)
Purpose
Format
Parameters
Description
klqbp The :CODE command should be sent only once before performing calibration.
Example

:COUNt?

(:CALibration:PROTected:CARD1:COUNt?
(:CALibration:PROTected:CARD2:COUNt?
Purpose
Format
Response
Description
klqb
Example
To program the calibration code or password so that you can perform the
Model 7751/7752/7753 calibration procedures.
:cal:prot:code '<char_string>'
Up to a 8-character string including letters and numbers.
The :CODE command enables the Model 2790 calibration procedures
when performing these procedures over the bus. This command must be
sent to the unit before sending any other Model 7751/7752/7753 calibra-
tion command. The default calibration code is KI002790.
Do not send :CODE before each calibration step.
The code parameter must be enclosed in single quotes.
:CAL:PROT:CODE 'KI002790'
To determine how many times a Model 7751/7752/7753 has been calibrated.
:cal:prot:card1:coun?
:cal:prot:card2:coun?
<n>
Calibration count.
The :CARD1:COUNt? and :CARD2:COUNt? commands allow you
to determine how many times a Model 7751/7752/7753 in Slot 1 or Slot 2
has been calibrated.
Use the :COUNt? command to help you monitor for unauthorized calibration
procedures.
:CAL:PROT:CARD1:COUN?
Send default code of KI002790.
Request card 1 calibration count.
G-17