Keithley SourceMeter 2790 Reference Manual page 170

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
Front panel scanning (7702 only)
When a simple scan using the 7702 module is configured, the present rel setting will apply
to all channels in the scan. When an advanced scan is configured, each channel can have
its own unique rel setting. Details to configure and run a scan are provided in
For an advanced scan, the following general procedure shows how to configure a scan
channel to use rel:
1.
2.
3.
4.
When the channel is scanned, rel will enable using the rel value established in step 3.
Remote scanning
For remote programming, the <clist> parameter is used to configure channels for a 7702
scan
allow scanning through 7751/7752/7753 channels (see
®
Switch System Reference Manual
While in the normal measurement state, select the appropriate function and close
the appropriate channel. For example, if scan channel 101 is going to be configured
for DCV and use rel, select DCV and close channel 101.
Apply the DCV signal to be rel'ed to the closed channel. This could be an offset or
a baseline level.
Press the REL key to enable rel (REL annunciator on). The input signal level is
used as the rel value.
When configuring the advanced scan (as explained in
channel, press DCV, and then press REL (REL annunciator on).
(Section
7). The <clist> parameter can also include memory pattern locations to
Section
Section
7), select the desired
"Memory patterns," page
5-3
7.
2-36).