Keithley SourceMeter 2790 Reference Manual page 139

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

3-34
Table 3-5 (cont.)
Basic measurement commands
Commands
CONT function
[SENSe[1]]
:CONTinuity:THReshold <NRf>
SYSTem:BEEPer <b>
Set temperature measurement units
UNIT:TEMPerature <name>
Trigger and retrieve readings
INITiate:CONTinuous <b>
INITiate[:IMMediate]
[SENSe[1]]
:DATA[:LATest]?
:DATA:FRESh?
FETCh?
READ?
Channel list parameter (7702):
<clist> = (@SCH)
where:
S = Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Channel list parameter (7751/7752/7753, DCV, Ω2, Ω4 only):
<clist> = (@Mn)
where:
M = Memory pattern command
n = Location number (1 to 40)
*Refer to the paragraphs following this table for more information.
1
The <clist> parameter is used for the module to configure one or more channels for a scan.
Model 2790 SourceMeter
1
Optional root command.
Set continuity threshold in ohms;
<NRf> = 1 to 1000.
Enable/disable beeper; <b> = ON or OFF.
Set temperature units; <name> = C, CEL,
F, FAR, or K.
Enable/disable continuous initiation;
<b> = ON or OFF.
Trigger one or more measurements.
Optional root command.
Returns the last reading string.
Returns the last "fresh" reading string.
Return reading(s).
Trigger and return reading(s).
®
Swich System Reference Manual
Description
Default Ref
10
ON
f*
g*
g*
g*
g*
g*