f.

Scanning programming example

The following program will scan 10 channels (101 through 110):
TRAC:CLE
INIT:CONT OFF
TRIG:SOUR IMM
TRIG:COUN 1
SAMP:COUN 10
ROUT:SCAN (@101:110)
ROUT:SCAN:TSO IMM
ROUT:SCAN:LSEL INT
READ?

Scanning examples

The following scanning examples assume that the Model 7702 switching module is
installed in slot 1 of the mainframe.
klqb
Tables are used for the procedure steps to configure and run scan examples. The left side
of the table provides the front panel procedure, while the right side shows the equivalent
remote programming commands. Where appropriate, menu sequences are provided to
summarize a front panel operation or selection. For example:
SHIFT SETUP > RESTORE: FACT
For the above menu sequence, press SHIFT and then SETUP to access the menu, use the
edit keys (
select it.
®
Switch System Reference Manual
Continuous initiation must be disabled in order to set the sample counter >1 (see
Reference c).
INITiate:CONTinuous, INITiate, and READ? — In order to initiate a single scan
cycle using INITiate or READ?, continuous initiation must be disabled. If you send
INIT or READ? with continuous initiation enabled, error -213 (Init ignored) will
occur.
You cannot use READ? or INITiate if sample count >1 AND there are readings
stored in the buffer by the TRAC command or by front panel data store operation
(error -225, out of memory). Either set the sample count to one or clear the buffer
(TRACe:CLEar).
A detailed example for 7751/7752/7753 memory pattern scanning is located in
Section
2. See
"Memory patterns" page
,
,
, and
' Clear buffer.
' Disable continuous initiation.
' Select the immediate control source.
' Set to perform one scan.
' Set to scan 10 channels.
' Set scan list channels; 101 through 110.
' Start scan when enabled and triggered.
' Enable scan.
' Trigger scan and request the readings.
2-36.
) to display RESTORE: FACT and then press ENTER to
7-33