Keithley SourceMeter 2790 Reference Manual page 152

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
Manual ranging
The range is selected by specifying the expected reading as an absolute value using the
<n> parameter for the appropriate :RANGe command. The Model 2790 will then go to the
most sensitive range for that expected reading. For example, if you expect a reading of
approximately 3V, let the parameter (<n>) equal 3 to select the 10V range.
Auto ranging
The :RANGe:AUTO command is coupled to the command to select range manually
(:RANGe <n>). When auto range is enabled, the parameter value for :RANGe <n>
changes to the automatically selected range value. When auto range is disabled, the instru-
ment remains at the selected range. When a valid :RANGe <n> command is sent, auto
ranging disables.
Range programming examples
Example #1 — The following commands select range for DCV, Ω2, and DCI:
VOLT:RANG 0.5
RES:RANG 2e3
CURR:RANG 0.1
Example #2 — The following command sequence configures channel 101 of the
Model 7702 to select the 10VDC range when it is scanned.
FUNC 'VOLT',(@101)
VOLT:RANG 1.5,(@101)
®
Switch System Reference Manual
' Select 1V range for DCV.
' Select 10kΩ range for Ω2.
' Select 100mA range for DCI.
' Set 101 for DCV function.
' Set 101 for 10V range.
4-5