15-14
Table 15-5 (cont.)
SENSe command summary
Command
[SENSe[1]]
:RESistance
:APERture <NRf>
[, <clist>]
:APERture? [<clist>]
:NPLCycles <n>[, <clist>]
:NPLCycles? [<clist>]
:RANGe
[:UPPer] <n> [, <clist>]
[:UPPer]? [<clist>]
:AUTO <b> [, <clist>]
:AUTO? [<clist>]
:DIGits <n> [, <clist>]
:DIGits? [<clist>]
:REFerence <n> [, <clist>]
:STATe <b> [, <clist>]
:STATe? [<clist>]
:ACQuire [, <clist>]
:REFerence? [<clist>]
:AVERage
:TCONtrol <name>
:TCONtrol?
:WINDow <NRf>
:WINDow?
:COUNt <n> [, <clist>]
:COUNt? [<clist>]
[:STATe] <b> [, <clist>]
[:STATe]? [<clist>]
7702 <clist> = (@SCH), where: S = Mainframe slot number (1 or 2) and CH = Switching module channel number.
7751/7752/7753 <clist> (DCV, Ω2, Ω4) = (@Mn) (see
Model 2790 SourceMeter
Description
Path to configure resistance.
Set integration rate in seconds (60Hz;
1.67e-4 to 1, 50Hz; 2e-4 to 1).
Query aperture integration rate.
Set integration rate in line cycles (60Hz;
0.01 to 60, 50Hz; 0.01 to 50).
Query line cycle integration rate.
Path to set measurement range:
Select range (0 to 120e6).
Query range.
Enable or disable auto range.
Query state of auto range.
Specify measurement resolution (4 to 7).
Query resolution.
Specify reference (0 to 120e6).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Path to configure and control filter:
Select filter type (MOVing or REPeat).
Query filter type.
Set filter window in % of range
(0 to 10).
Query filter window.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
"Memory patterns," page
®
Switch System Reference Manual
Default
parameter
(Note 2)
5.0
120e6
ON
7
0
OFF
(Note 3)
0.1
10
(Note 4)
2-36).
Ref
SCPI
Sec 3
Sec 4
Sec 4
Sec 4
Sec 4
Sec 5
Sec 4