C-4
OComp (offset-compensated ohms)
The Model 2790 performs a normal ohms measurement by sourcing a known current (I),
measuring the voltage (V), and then calculating the resistance (R = V/I). Offset-
compensated ohms cancels the effects of thermal EMFs which can adversely affect low-
resistance measurements.
With OCOMP enabled, the Model 2790 performs one normal resistance measurement and
then loops back to perform a second resistance measurement with the internal current
source set to its lowest level. The offset-compensated ohms reading is then calculated as
shown in
klqb
Filter
The filter is used to stabilize noisy readings. With the filter enabled, the specified number
of readings are averaged to yield a single filtered reading. There are two types of filters:
moving and repeating.
A filter stack is used to temporarily store the specified number of readings to be averaged.
In general, for the moving filter, each measurement process adds a reading to the stack
(oldest reading discarded), and then averages the stack to yield a filtered reading. For the
repeating filter, each measurement process fills the stack with new readings (all previous
readings discarded), and then averages the stack to yield a filtered reading.
klqb
Output trigger pulse (VMC)
An output trigger pulse from the Model 2790 can be used to trigger an external instrument
to perform an operation. In general, a trigger pulse is output at this point in flow chart for
each processed reading.
An exception is the SCAN function for scanning. For the SCAN function, an output trig-
ger is not output until after the specified number of channels (as set by the sample counter)
are scanned.
klqb
Rel
Next in the signal processing sequence is the Rel operation. Rel is used to null offsets or
subtract a baseline rel value from the reading. With Rel enabled, the Rel'ed reading is
calculated as shown in
klqb
Model 2790 SourceMeter
Figure
C-2.
For details on OCOMP measurements,
page
3-20.
For details on filter operation, see
For details on scan operation, see
Figure
C-2.
For details on Rel operation, see
®
Switch System Reference Manaul
see"Offset-compensated ohms,"
"Filter," page
4-13.
"Trigger models," page
"Relative," page
5-2.