Remote Programing - Limits And Digital Output; Limits And Digital Output Commands - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter Switch System Reference Manual
Remote programing — limits and digital output

Limits and digital output commands

The limits and digital output commands are provided in
Table 9-2
Limits and digital I/O commands
Commands
Limit 1 commands
CALCulate3:LIMit1:UPPer <NRf>
[, <clist>]
CALCulate3:LIMit1:LOWer <NRf>
[, <clist>]
CALCulate3:LIMit1:STATe <b> [, <clist>] Enable/disable Limit 1 test; <b> = ON or OFF.
CALCulate3:LIMit1:FAIL?
CALCulate3:LIMit1:CLEar
CALCulate3:LIMit1:CLEar:AUTO <b>
Limit 2 commands
CALCulate3:LIMit2:UPPer <NRf>
[, <clist>]
CALCulate3:LIMit2:LOWer <NRf>
[, <clist>]
CALCulate3:LIMit2:STATe <b>
[, <clist>]
CALCulate3:LIMit2:FAIL?
CALCulate3:LIMit2:CLEar
CALCulate3:LIMit2:CLEar:AUTO <b>
*
Set HI1 limit; <NRf> = -4294967295 to
+4294967295.
Set LO1 limit; <NRf> = -4294967295 to
+4294967295.
Query test result; 0 = pass (in), 1 = fail
(high or low).
Clear fail indication.
Enable/disable auto clear; <b> = ON or OFF.
Set HI2 limit; <NRf> = -4294967295 to
+4294967295.
Set LO2 limit; <NRf> = -4294967295 to
+4294967295.
Enable/disable Limit 2 test; <b> = ON or OFF.
Query test result; 0 = pass (in), 1 = fail
(high or low).
Clear fail indication.
Enable/disable auto clear; <b> = ON or OFF.
Table
9-2.
Description
9-13
Def
Ref
1
-1
a
b
c
ON
c
2
-2
a
b
c
ON
c