Keithley SourceMeter 2790 Reference Manual page 256

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
The Trigger Link connections for this test system are shown in
the Model 2790 is connected to Trigger Link (either IN or OUT) of the Model 7002. Note
that with the default trigger settings on the Model 7002, line #1 is an input and line #2 is
an output. This complements the trigger lines on the Model 2790.
For this example, the Models 2790 and 7002 are configured as follows:
Model 2790
Factory defaults restored (accessed from SHIFT-SETUP)
External triggers (accessed from EX TRIG)
Buffer enabled and set to store 400 readings
Model 7002
Factory defaults restored
Scan list = 1!1-1!400
Number of scans = 1
Channel spacing = TrigLink
Figure 8-7
Trigger link connections
WARNING:
INTERCONNECTION, INST ALLA TION AND REMOV AL OF CARDS BY QUALIFIED SERVICE PERSONNEL ONL Y .
CARD
CARD
CARD
1
2
3
WARNING:
NO INTERNAL OPERA TOR SERVICEABLE P ARTS, SERVICE BY QUALIFIED PERSONNEL ONL Y .
®
Switch System Reference Manual
Model 7002
CARD
CARD
CARD
CARD
4
5
6
7
Trigger Link
CARD
CARD
CARD
8
9
10
TRIGGER
LINK
DIGIT AL
I/O
A1
A2
A3
A4
A5
CHANNEL
READY
EXTERNAL
IEEE-
TRIGGER
MADE IN USA
Trigger Link
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
DIGITAL I/O
TRIG. LINK
RS232
MADE IN
!
KEITHLEY
SLOT COVER
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Model 2790
Figure
8-7. Trigger Link of
I
N
O
U
T
488
Trigger Link
LINE
R A T I N G
90-260
V
Cable (8501)
47-440
Hz
85
V A
MAX
IEEE-488
U.S.A.
SLT
1
SLT
2
8-11
!

Advertisement