Model 7751, 7752, and 7753 calibration commands

Table G-4
modules. Note that CARD1 commands calibrate the card in Slot 1, while CARD2 com-
mands request calibration count and date information from a card in Slot 2.
klqb
Table G-4
Model 7751/7752/7753 calibration commands
Command
:CALibration
:PROTected
:CODE '<up to 8 char. string>' Send calibration code. (Default KI002790.)
:CARD1
:INITiate
:COUNt?
:RCOunt
:DATE?
:SAVE
:LOCK
:LOCK?
:ISOurce
:STEP0
:STEP1
:DATA <NRf>, <NRf>
:DATA?
:CARD2
:COUNt?
:DATE?
Model 2790 SourceMeter® Switch System Reference Manual
summarizes calibration commands for the Model 7751, 7752, and 7753 plug-in
A Model 7751/7752/7753 must be installed in Slot 1 to be calibrated. See
"Model 7751, 7752, and 7753
Calibration root command.
All commands in this subsystem are protected by the calibration
lock (except queries and :CODE).
Path to calibrate card in slot 1.
Initiate card calibration.
Request number of times card in slot 1 has been calibrated.
Reset card calibration count to 0.
Request card cal. date.
Save cal constants to card EEPROM.
Lock out calibration.
Request cal lock state. (0 = locked, 1 = unlocked.)
Path to Model 7751/7752/7753 current source commands.
Turn on current source low calibration point (1mA).
Turn on voltage source high calibration point (50mA).
Send two DMM readings for low and high calibration points.
Return comma-delimited calibrations constants.
Path to request calibration information from card in slot 2.
Request number of times card in slot 2 has been calibrated.
Request card calibration date.