Manage Custom Methods
Create Custom Method
Thermo Scientific
The NanoDrop One PC Control software is your tool for creating and managing
custom methods, which contain user-defined settings that can be used to acquire
data with the instrument. Custom methods can be made with or without standards.
Create method to be used for sample measurements with user-defined settings.
Create new custom method
–
From the NanoDrop Home screen, select the Custom tab and select
Custom Methods
Custom
Tab
Custom
Method
Icon
–
In the Manage Custom Methods screen, select
one of the following:
–
Formula (if your method will not have standards)
–
Standard Curve (if your method will have standards)
–
In the setup window, enter Method Name (this name appears in the
Custom Setup
there)
–
Enter detailed Description of method, if desired
–
Specify how to calculate and report the method results:
–
if method does not have standards, specify
coefficient of analyte
only)
–
if method has standards, enter
standard and select the curve fit type
box on the instrument after the method has been transferred
(enter "1" to report absorbance measurements
7
Custom Applications
factor or extinction
name and concentration of each
NanoDrop One User Guide
Measure Custom
and choose
159