Setting
Description
Curve Type
Specify type of equation used to create standard curve from standard concentration
values. Available options:
Replicates
Enter number of measurements of the reference or the same standard or sample
that are averaged together to produce its associated concentration value.
Note: Replicates setting cannot be changed after the first standard has been
measured.
Standards
Enter actual concentration value of each standard.
Note: Concentration values can be entered in any order but the standards must be
measured in the order they were entered.
Thermo Scientific
–
Linear: Draws the linear least squares line through all measured standards
(requires reference measurement and at least one standard)
–
Interpolation: Draws a series of straight lines to connect all measured
standards (requires reference measurement and at least one standard)
nd
–
2
order polynomial: Draws the 2
using all measured standards (requires reference measurement and at least
two standards)
rd
–
3
order polynomial: Draws the 3
all measured standards (requires reference measurement and at least three
standards)
Related Topics
•
Instrument Settings
5
Protein Applications
Measure Protein Lowry
nd
order least squares polynomial
rd
order least squares polynomial using
NanoDrop One User Guide
125