3. Measure a blank:
–
Pipette 1–2 µL blanking solution onto the lower
pedestal and quickly lower the arm
–
Tap Blank and wait for the measurement to
complete
Tip: If
Auto-Blank
starts automatically after you lower the arm.
–
Lift the arm and clean both pedestals with a new
laboratory wipe
4. Measure the first sample:
–
Pipette 1-2 µL sample solution onto the pedestal
and quickly lower the arm (see
Sample Volumes
–
Start the sample measurement:
–
if
Auto-Measure
–
if Auto-Measure is off, lower arm and tap
Measure
–
When the sample measurement is completed, the
spectra and reported values are displayed.
5. To measure another sample:
–
Lift the arm
–
Clean both pedestals with new wipe
–
Load the next sample and quickly lower the arm
–
Start the sample measurement
–
Wait for the measurement to complete
The new spectrum replaces the previous one on the
spectral display and the new reported values appear
under the previous ones in the table. (Drag tab down
to show both sets of data.)