5
Protein Applications
Measure Protein A280
70
NanoDrop One User Guide
• From the local control Home screen, tap
from the Protein A280 or Proteins & Labels measurement screen, tap
Settings > Protein Editor.
Local Control Protein Editor
Tap to add custom protein
Custom proteins (will appear in Sample
Type list in Protein A280 Setup and
Proteins & Labels Setup)
Tap to close Protein Editor
These operations are available from the Protein Editor:
Add custom protein
1. In Protein Editor, select
box.
2. Enter a unique Name for the new protein (from local control, tap field to display
keyboard, tap Done key to close keyboard).
3. Enter a Description for the new protein.
4. Specify whether to enter Molar Extinction coefficient or Mass Extinction
coefficient for custom protein.
If Mass Extinction coefficient is selected, enter mass extinction
coefficient in L/gm-cm for 10 mg/mL (
> Protein Editor. Alternatively,
or
to show the New Protein Type
1%) protein solution.
>
Tap to edit
selected custom
protein
Tap to delete
selected custom
protein