Operating Specifications
Thermo Scientific
The instrument operates reliably when the room environment meets these
specifications:
• operating temperatures: 5 °C - 35 °C (41 °F - 95 °F)
• relative humidity (non-condensing): 20-80%
Locate the instrument away from air vents and exhaust fans to minimize evaporation.
Note If operating the instrument at the low end of the recommended humidity
range, use adequate sample volume to avoid evaporation.
After the instrument is installed, you can leave it turned on.
9
Learning Center
Operating Specifications
NanoDrop One User Guide
195