lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
repeat substeps above for each additional standard (when specified number
of standards and replicates have been measured, a message asks whether
to load more standards or begin measuring samples)
if finished measuring standards, tap Done (swipe left to view standard curve)
6. Measure samples:
pipette 2 µL sample 1 onto pedestal, or insert sample 1 cuvette
lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
7. When you are finished measuring samples, tap End Experiment.
8. Lift the arm and clean both pedestals with a new wipe, or remove the sample
cuvette.
Related Topics
• Working with standard curves
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Prepare Samples and Blanks
• Basic Instrument Operations
5
Protein Applications
Measure Protein Lowry
NanoDrop One User Guide
121