Maintenance

Maintaining the Cuvette Sampling System

Maintaining the Cuvette Sampling System

To maintain the cuvette sampling system
• Keep the instrument arm closed when the
instrument is not in use.
• Use canned air to remove any dust from the
cuvette holder.
• Clean up any spills inside the cuvette holder
with a new laboratory wipe.
To clean and maintain cuvettes, follow the
recommendations of the cuvette manufacturer.

Instrument Diagnostics

260
NanoDrop One User Guide
The cuvette sampling system is included only with the NanoDrop One
instrument. For information about compatible cuvettes, see
using a
Cuvette.
Note Clean and dry cuvettes after each measurement. Use cuvettes that are
free of scratches and avoid fingerprints which may affect results.
Note Do not use a squirt or spray bottle on or near the instrument as liquids will
flow into the instrument and may cause permanent damage.
Every 6 months, run the following performance and quality checks to verify
instrument operation.
Intensity Check 261
Performance Verification 263
Pedestal Image Check 268