Thermo Scientific
if Molar Extinction is selected,
enter molar extinction coefficient (
is,
/1000). For example, for protein with molar extinction coefficient
of 210,000 M
Enter molecular weight (MW) in kiloDaltons (kDa)
5. From local control, tap OK to close the New Protein Type box. From PC
control software, select Save.
6. Enter your password to sign the changes if prompted.
The new custom protein appears in the Type list in Protein A280 Setup and
Proteins & Labels Setup.
Tap a field to show keyboard; to
close, tap Done key
) in M
-1
-1
cm
, enter 210.
5
Protein Applications
Measure Protein A280
-1
-1
cm
divided by 1000 (that
Enter your
password to
sign the
changes
NanoDrop One User Guide