To measure OD600 samples

Thermo Scientific
Measure a series of dilutions using a young overnight culture (~16 hrs) of the
microbial strain
Graph the OD600 measurements against the dilution factor
The upper detection limit is the measured OD600 value at which there ceases to
be a linear correlation between dilution factors and OD600 readings.
• Mix samples gently but thoroughly immediately before taking an aliquot for
measurement.
• For micro-volume measurements:
Ensure pedestal surfaces are properly
Avoid introducing bubbles when mixing and pipetting.
Start the measurement promptly to avoid settling or evaporation.
Follow
best practices for micro-volume measurements.
Use 2 µL sample volume. See
information.
For dilute samples that exhibit low absorbance at 600 nm, use an alternative
wavelength such as 400 nm to measure absorbance, or use cuvettes
instead of micro-volume measurements.
• For cuvette measurements (NanoDrop One
Use clean plastic, glass or quartz cuvettes.
Follow
best practices for cuvette measurements.
Do not use the automatic
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will
flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will
permanently damage the quartz fiber optic cables.
cleaned
and
Recommended Sample Volumes
C
instruments only):
stirring
feature for this assay.
NanoDrop One User Guide
6
Measure OD600
conditioned.