Thermo Scientific NanoDrop One User Manual page 104

Micro-uv/vis spectrophotometer
Hide thumbs Also See for NanoDrop One:
Table of Contents

Advertisement

5
Protein Applications
Measure Protein BCA
102
NanoDrop One User Guide
Tip: If using a cuvette, make sure to
light path.
tap Blank and wait for measurement to complete
lift arm and clean both pedestals with new laboratory wipe, or remove
cuvette
4. Measure reference standard:
pipette 2 µL reference solution onto pedestal, or insert reference cuvette
(reference solution should contain none of the standard protein stock, see
Working With Standard Curves
lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
5. Measure remaining standards:
pipette 2 µL standard 1 onto pedestal, or insert standard 1 cuvette
lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
repeat substeps above for each additional standard (when specified
number of standards and replicates have been measured, a message asks
whether to load more standards or begin measuring samples)
if finished measuring standards, tap Done (swipe left to view standard
curve)
6. Measure samples:
pipette 2 µL sample 1 onto pedestal, or insert sample 1 cuvette
lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
7. When you are finished measuring samples, tap End Experiment.
8. Lift the arm and clean both pedestals with a new wipe, or remove the sample
cuvette.