Advertisement

5
Protein Applications
Measure Protein BCA
Protein BCA standard curve screen
104
NanoDrop One User Guide
Note
• A baseline correction is performed at 750 nm (absorbance value at 750 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements and measurements taken with
nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
The standard curve screen shows graphically the relationship between the
measured standards, the calculated standard curve, and the measured absorbance
and calculated concentration for a selected sample. A horizontal line connects the
sample absorbance value on the Y-axis to the standard curve. A vertical line
connects that point to the sample concentration value on the X-axis.