Thermo Scientific NanoDrop One User Manual page 267

Micro-uv/vis spectrophotometer
Hide thumbs Also See for NanoDrop One:
Table of Contents

Advertisement

Thermo Scientific
4. Pipette 1 µL DI H
O onto the lower pedestal, lower the arm and tap Blank.
2
5. Lift the arm and clean both pedestals with a new wipe.
Note Vigorously shake the ampoule of PV-1 solution, allow the liquid to
collect in the bottom portion of the ampoule and then follow standard
practices to open it.
6. Pipette 1 µL PV-1 solution onto the lower pedestal and start the sample
measurement:
• If
Auto-Measure
• If Auto-Measure is off, lower arm and tap Measure or from the PC Control
software, select Run.
After the measurement, the software displays the results. Here is an example of
the performance verification result screen.
7. Repeat step 6 to measure the PV-1 solution nine more times using a new
1 µL aliquot for each measurement and cleaning both pedestals after each
measurement.
is On, lower arm
Measured pathlengths
Drag tab down/up to see
more/less data
Swipe screen left to view
detailed results
10
Maintenance
Instrument Diagnostics
NanoDrop One User Guide
265