SETup Subsystem

Description

The SETup subsystem is used to configure the test set for each measurement. Typical settings include:
• Multi-Measurement Count, how many measurements will be made each time a measurement is initiated.
• Trigger Arm, determines if a measurement will make one measurement then return to idle (single), or
automatically rearm on completion of a measurement and repeat the process (continuous).
• Trigger Source, how a measurement is triggered.
• Trigger Delay, controls the delay time between the trigger and the start of sampling.
• Measurement Timeout, length of time before a measurement times out.
Trigger arming for each measurement is controlled in the SETup subsystem. The choices are
NOTE
single or continuous. In most cases, it is a best practice (during remote operation) to use "single"
measurement mode. This simplifies the tasks of starting concurrent measurements, using the
INIT subsystem commands to determine which measurements are ready to be fetched, then
using the FETCh subsystem to obtain results. The command "SETup:CONTinuous:OFF sets all
measurements to "single" trigger mode.

Syntax Diagrams and Command Descriptions

"SETup:AFANalyzer" on page 285
"SETup:WACLeakage" on page 294
"SETup:WBERror" on page 300
"SETup:WCPower" on page 303
"SETup:WTPower" on page 308
"SETup:WWQuality" on page 312