CALL:STATus:AWGNoise:POWer[:AMPLitude]?
Function
Queries the current power level of the additive white Gaussian noise (AWGN) source.
AWGN power level is set using the
Query
Range:
• -170dBm/3.84 MHz to +37dBm/3.84 MHz, 9.91 E+37
(-120 dBm/3.84 MHz to -13 dBm/3.84 MHz actual hardware range with amplitude offset = 0)
The actual power range is defined by adding the value of the associated amplitude offset to
the range above. For example, if amplitude offset is -3.5 dB, then the power range will be
-123.5 dBm/3.84 MHz to -16.5 dBm/3.84 MHz.
Resolution: 0.01 dBm/3.84 MHz
*RST Setting
9.91 E+37
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:AWGNOISE:POWER:AMPLITUDE?"
!Queries the current AWGN power.
CALL:STATus:AWGNoise:POWer:STATe?
Function
Queries the current on/off state of the additive white Gaussian noise (AWGN) source.
AWGN power level on/off state is set using the
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:AWGNOISE:POWER:STATE?"
"CALL:AWGNoise:POWer"
"CALL:AWGNoise:POWer"
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_call_status.fm
commands.
commands.
CALL:STATus