CALL[:CELL]:POWer
CALL[:CELL]:POWer[:SAMPlitude][:SELected]
Function
The setting form of this command performs two functions: (1) sets the cell power level and (2)
turns the cell power on/off state to on. The query form returns the cell power level (including
compensation for amplitude offset) or 9.91 E+37.
The optional [:SELected] keyword in this command specifies that the cell power level being set or
queried applies to the current system type. The current system type can be changed using the
"CALL[:CELL]:SYSTem[:TYPE]?"
Setting
Range:
• -170dBm/3.84 MHz to +37dBm/3.84 MHz, 9.91 E+37
(-120 dBm/3.84 MHz to -13 dBm/3.84 MHz actual hardware range with amplitude offset = 0)
The actual power range is defined by adding the value of the associated amplitude offset to
the range above. For example, if amplitude offset is -3.5 dB, then the power range will be
-123.5 dBm/3.84 MHz to -16.5 dBm/3.84 MHz.
Resolution: 0.01 dBm/3.84 MHz
Query
Range: Same as setting range.
Resolution: 0.01 dB
*RST Setting
9.91 E+37
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:CELL:POWER:SAMPLITUDE -30"
!Sets the cell power state to ON and sets RF output power to -30 dBm.