SYSTem:PRESet

:PRESet2
SYSTem
:PRESet3
"Diagram Conventions" on page 168
SYSTem:PRESet2
Function
Performs a full preset of the test set. This is the remote equivalent of pressing the SHIFT Preset
keys on the front panel of the test set.
All parameters are set to their default values. All measurements are aborted the trigger arm is
set to continuous. See
The *RST command will set the trigger arm to single.
Related Topics
See
"Full Preset" on page 117
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SYSTEM:PRESET2" !Full preset, trigger arm set to continuous.
SYSTem:PRESet3
Function
Performs a partial preset. This is the recommended command for users when a partial preset
is needed during remote operation of the test set.
Any call in process is disconnected and all measurements are aborted and inactivated.
Measurement parameters are not changed.
A partial preset will not modify any measurement settings including trigger arm.
Related Topics
See
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SYSTEM:PRESET3" !Partial preset when in remote operation.
( full preset trigger arm continuous)
(partial preset trigger arm no change)
"Trigger Arm (Single or Continuous) Description" on page
for details about the *RST full preset.
"Partial Preset"
for more details.
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_system_preset.fm
SYSTem:PRESet
150.