Error Message
+460 Hardware failure; Hardware doesn't support
requested operation
+461 Hardware failure; Link subsystem hardware is not
responding. Call processing functionality not operational.
+462 Hardware error; Hardware not supported; <board
ID><part_number>
+463 Hardware error: RFIO version does not support
loopback switching.
+464 Hardware error: RFIO version does not support RF
out only selection.
<board ID> names
3 GHZ ATTENUATOR 1| ;3GHZ ATTENUATOR 2| ;AUDIO BD| ;ROM BASEBAND GENERATOR 1| ;ROM
BASEBAND GENERATOR 2| ;DEMOD DOWNCONVERTER| ;VECTOR OUTPUT BOARD 1| ;VECTOR
OUTPUT BOARD 2| ;IVF MEASUREMENT| ;MEASUREMENT DOWCONVERTER| ;RF POWER
DETECTORS| ;REFERENCE MODULE| ;SYNTH DOUBLER 1| ;SYNTH DOUBLER 2| ;TIMING REF|
;MOMENTUM INSTRUMENT| ;RF MOTHER BOARD| ;JUMPER BOARD| ;DIGITAL MOTHER BOARD|
;FLAT PANEL ADAPTER| ;REAR PANEL BOARD
<EEPROM board ID> names
Instrument Eeprom ID State| ;Atten 1 Eeprom ID State| ;Atten 2 Eeprom ID State| ;Audio Eeprom ID
State| ;BaseBandGen 1 Eeprom ID State| ;BaseBandGen 2 Eeprom ID State| ;Demod DC Eeprom ID State|
;Digital Mother Board Eeprom ID State| ;IQ Output 1 Eeprom ID State| ;IQ Output 2 Eeprom ID State| ;IVF
Meas Eeprom ID State| ; Jumper Board Eeprom ID State| ; Meas DC Eeprom ID State| ;RF Mother Board
Eeprom ID State| ;RF Interface Eeprom ID State| ; Ref Mod Eeprom ID State| ;Sig Gen 1 Eeprom ID State|
; Sig Gen 2 Eeprom ID State| ;Time Ref Eeprom ID State| ;Display Interface Eeprom ID State| ;Rear Panel
Eeprom ID State
Description
This error occurs when a setting is attempted that
is not possible with the test set's hardware. The
current firmware needs new hardware in order to
work.
This error occurs any time the link subsystem is not
operating correctly. The message is persistent to
warn the user that any call processing operations
will not work.
This error appears when an operation was
attempted that requires hardware that is not
installed in the test set.
This error occurs when trying to select the RF OUT
ONLY port when the test set has the RF I/O module
that only has the RF IN/OUT port.